I. Brief Introduction:
1. The battery has been changed to a rechargeable battery, avoiding the hassle of replacing batteries and the faults caused by it.
2. The voltage is divided into three adjustable levels to accommodate counters of different models and brands.
3. The packaging has been improved to an aluminum alloy case, which is sturdy and durable for consumer use.
II. Technical Parameters
1. Voltage: DC 900V, 1200V, 1600V three levels
2. Discharge current >100A/10US
III. Usage Instructions
1. Charge before use. To ensure charging effectiveness and prolong battery life, use the dedicated charger. It takes 3-4 hours to fully charge an empty battery. A red light on the charger indicates charging; a green light indicates charging is complete. Before use, connect the attached ground wire plug at the rear end, and clamp the ground clip to the earth.
2. According to the installation height of the counter, properly extend the telescopic discharge rod.
3. Press the red button to apply high voltage for about 1 second. When the indicator light turns on (flickering), gently tap the connection end between the counter and the arrester for testing.
4. After each tap, the discharge rod tip should be removed from the counter. If repeated testing is needed, you do not need to release the button. After about 1-2 seconds, when the indicator light flickers again, you can tap again for testing.
5. Continuous testing may cause the calibration instrument to heat up; please allow appropriate intervals.
6. If the indicator light does not flicker within 3 seconds of pressing the button, the battery needs to be charged.
7. Do not disassemble the calibration instrument arbitrarily. If the battery capacity significantly drops or charging efficiency is too low, it needs replacement. Please purchase the dedicated battery pack from our company.
Note:When the indicator light is on(flickering), no part of the body should touch the electrode part,otherwise there is a risk of electric shock.